Appendix 5b: pH Meters

pH meters are electronic devices which measure the pH of an
aqueous solution. We will cover the
theory behind pH meters when we study electrochemical cells. Let it suffice for now that the pH probe is
a combination electrode containing two half cells, the pH half cell and a
reference half cell. The potential
across these cells is logarithmically related to the hydronium ion
concentration (see Nernst eq.) and thus a measure of the pH. For this reason the pH meter has two scales,
a pH scale and a voltage scale.
Before making any measurements you need to calibrate the pH
meter. This is done by successively
placing the electrodes into different buffer solutions of increasing pH while
the instrument is operating in calibration mode. When you accept each calibration point the meter locks that value
in. If you run the meter for several hours
you should redo a calibration to offset any electrode potential drift.

Calibration Techniques:
We will use 3 buffers of pH
4.00
7.00
10.00
Pour about an inch of each buffer
into a separate beaker.

Make sure the solutions are
deep enough to cover the bottom of the
electrode. Do not
waste buffer solutions
by using excess buffer.
Fill a
beaker with distilled water and
wash. Immerse the
electrode in the
water and then wash again with a
more distilled water in a squeeze bottle.

In order to prevent a dilution effect
as you make your measurements
you should dry the outside of the
electrode with a Kim wipe prior to
immersing in any solutions.
Never dry the electrode itself,
just the outside covering.

Place the electrode
into the most acidic buffer.

If the pH meter does not read
in pH, press mode until it does
-the arrow on the bottom of the
display points to pH
and not mV or Rel
mV
-Press "2nd Calibrate"
The meter now displays calibrate
above the
reading and states
P1 below (indicating this is the
first calibration value.)
When it reads "ready", hit "yes" if you
want to
automatically accept the value and "no"
if you want to manually set the value. If you
hit no, you now have to hit the up or down
button twice and the left most digit will change.
Once it is correct, hit "yes"
and you will be toggled to the next digits position.
Repeat
until finished.
Now the meter will change to P2 and you
are ready to take the second reading.
Don't forget to wash the probe
between successive measurements
When you accept the second value
the meter will prompt you for a third
value (pH
10) if it uses a three point
calibration or it will display the slope
if it is uses a 2 point calibration.
The buffers need to bracket the pH range
you desire to test.
(The above values would
be inappropriate to measure the pH of a
basic solution.)
The value of the slope should be
between 95
and 105.
The meter is now calibrated and you are
ready to run your experiments.
Note how it reads "measure" and
not "calibrate"
"ATC" means automatic temperature
compensation. This
allows you to measure
samples at different temperatures than the
standards used in the calibration.
A final note, never leave the probes outside or let them
dry. They come in a test tube
containing a saturated potassium chloride solution. It is OK if some of the solution evaporates and KCl crystals form
on the outside of the test tube as long as the electrode is immersed in
solution. Notify your instructor if
your electrodes are not immersed in solution during the beginning of your
experiment.